pH meter input impedance and its measurement method

<

1. Input Impedance When a pH meter is used to measure the acidity of a solution, a glass electrode and a calomel electrode constitute a chemical primary battery in a solution. It has an electromotive force E and an internal resistance r. Therefore, the input impedance of the pH meter and the internal resistance of the primary battery can be equivalent.

E represents the electromotive force of the primary battery, and its value is related to the pH value of the measured solution; r represents the internal resistance of the primary battery, which consists of 3 parts, mainly determined by the internal resistance of the pH meter glass electrode (about 108Ω); The internal resistance of mercury electrode is about 104Ω; the internal resistance of the tested solution is (103~105)Ω. R represents the input impedance of the pH meter, which is the parallel value of the resistance of each part of the input of the pH meter.

The internal battery resistance r is in series with the input impedance R. According to the principle of series resistance voltage division, when the internal resistance of the primary battery is 109Ω, the input impedance of the pH meter should be more than 1000 times larger than the internal resistance of the primary battery, which is more than 1012Ω. Therefore, the influence of the voltage drop on the primary battery r can be ignored, so that the voltage entering the input terminal of the pH meter is close to the electromotive force of the primary battery.

When the input impedance of the pH meter is not large enough, if the input impedance is equal to the internal resistance of the primary battery, there will be half the pressure drop in the primary battery. The value displayed on the pH meter is only half of the electromotive force of the primary battery. Even if the input impedance of the pH meter is one to two orders of magnitude larger than the internal resistance of the primary battery, instability will occur during measurement. This is because the input impedance of the pH meter and the internal resistance of the galvanic cell are not all completely stable constants; they vary with the ambient temperature and humidity.

The input impedance of the so-called pH meter is the impedance seen from the two inputs of the pH meter. The input impedance of the meter is not only related to the high-impedance tube at its input, but also related to the reading of the input switch, the glass electrode jack, the insulation resistance of the input shielded wire, and the filter capacitor leakage resistance at the input. Because, in principle, their insulation resistance is in parallel with the input high-impedance tube. If the reading switch and the glass electrode socket are installed in the pH meter casing, there is a leakage resistance between them and the casing. The voltage signal on the electrode enters the input of the high-impedance tube through the shielded line, and the input filter capacitor is also grounded. therefore. Their insulation resistance or leakage resistance is at least two orders of magnitude greater than the impedance of a high-impedance tube. If one of the above components has an insulation resistance that does not meet the requirements, it will affect the input impedance of the pH meter. If the above components are contaminated, they must be cleaned. The cleaning solvent uses diethyl ether instead of ethanol. After cleaning, it is dried with a hair dryer. The insulation resistance of the components after cleaning can generally meet the requirements.

Second, the pH meter input impedance measurement method The input impedance of the pH meter can not be measured directly, can be obtained by indirect measurement.

R Take 1000MΩ Input voltage E0 from potentiometer to pH meter When switch K is on (R is short-circuited), millivolt value measured with pH meter is E0; switch K is then turned off and R is turned on The millivolt value obtained is Ei, the following formula can be obtained:

Ei=(Ri×E0)/(R Ri)(1)

Where: Ri - the input resistance of the pH meter.

From equation (1), Ri = (R × Ei) / (E0 - Ei) (2)

Third, the calculation example Select a 0.01 pH meter, according to Figure 2 wiring. R is 1000MΩ, the input voltage is 300mV from the potentiometer to the pH meter, and the value of the pH meter is 300.0mV when the switch K is turned on (R is short-circuited); then the switch K is turned off (R is turned on), and the pH meter is turned off. The indicated value is 299.8mV. Substituting the measured data into equation (2) yields Ri=(R×Ei)/(E0-Ei), Ri=1.50×1012Ω

1. Honing pipe is essentially a specially processed high precision seamless steel tube, which has good performance in all fields.

2. the company of Honing Tube basic process: (1) the material are all made of the production of TPCO tianjin big seamless continuous rolling pipe, (2) tube material inspection, pickling, phosphating, saponification (3) cold drawing processing form Cold Drawn Tube (4) the cold drawing for honing honing machine.   

  Honing is also a way of polishing. Honing is only for the polishing of holes in the steel tube. Honing is a high-precision grinding process. Honing has a dedicated honing head, which is equipped with various thick oil and stone to polish the inner surface of the inner hole. It is installed on the floating clamping device, which can self-regulate in a small amount, so it can adjust the shape error of the hole. Its working state is that the oil stone makes slow circular motion and relatively fast axial movement, making the inner hole size meet the process requirements.

3. The process and characteristics of honing steel tube and improving surface roughness can almost reach Ra less than 0.4 um. 2. Correct the roundness, the ellipticity can be less than 0.01 mm. 3, improve the surface hardness, eliminate the stress deformation, hardness HV 4 ° or higher

4. After processing, the residual stress layer shall improve the fatigue strength by 30%.

5. Improve the coordination quality, reduce wear and tear, and prolong the life of cylinder.

Internal diameter tolerance: H8, H9, straight line 0.3mm/ 1000mm, 0.5mm/ 1000mm.

 

The company adopts the 250T hydraulic cold-drawing machine for cold drawing, the inner diameter of product specification: 40mm- 630mm, outer diameter 50mm- 820mm, wall thickness: 5cm-40cm, cold-drawn tube single branch length is less than 13.5 meters

 

Steel tube weight calculation formula:

[(outer diameter - wall thickness) * wall thickness] *0.02466= kg/meter (weight per meter)

Honing Tube

Honing Tube,Seamless Honing Tube,Carbon Hydraulic Honed Tube,Honed Steel Tube

Yangzhou Yinhe Machinery Co., Ltd , http://www.yinhesteeltubes.com